Recalls / —
—#214408
Product
Dual Flat Panel Monitor (DFPM) accessories for HarmonyAIR A-Series Surgical Lighting Systems
- FDA product code
- FSY — Light, Surgical, Ceiling Mounted
- Device class
- Class 2
- Medical specialty
- General, Plastic Surgery
- Affected lot / code info
- UDI: 00724995203245: Serial Numbers: 0400224028, 0401625167, 0401625171, 0401625180, 0401625185, 0401625190, 0402022165, 0402022169, 0402422003, 0402622007, 0402925148, 0403122053, 0403122057, 0403125009, 0403223079, 0403223083, 0403223087, 0403223091, 0403321066,0403321070, 0403423063, 0403423067, 0403423071, 0403423075, 0403423079, 0403423083, 0403423087, 0404022003, 0404622011, 0404622014, 0404622017, 0404622020, 0404622023, 0404822104, 0405025032, 0405924100, 0405924103, 0406624134, 0406624138, 0406624142, 0406722052, 0406722056, 0406722060, 0406722064, 0406823005, 0407625022, 0408025008, 0408222120, 0408222124, 0408222128, 0408222132, 0408922093, 0408922097, 0408922101, 0409622008, 0409822014, 0409822018, 0410822042, 0410822046, 0410823068, 0410923149, 0410923152, 0410923155, 0413223057, 0413223060, 0414223170, 0414223174, 0414223178, 0414223182, 0414223186, 0414223190, 0414223194, 0414223198, 0414323003, 0414323006, 0415223065, 0415223068, 0415223071, 0415223074, 0415923036, 0416924069, 0416924074, 0417422053, 0418023091, 0418023095, 0418023099, 0418023103, 0418023107, 0418023111, 0419223082, 0419523010, 0419724177, 0419924007, 0420223048, 0420223051, 0420223054, 0420223057, 0420223060, 0420823170, 0421322041, 0421322045, 0421623206, 0421623211, 0421623216, 0421623221, 0421623226, 0421623231, 0421623236, 0421623241, 0422923010, 0423023059, 0423023304, 0423023307, 0423023379, 0423023384, 0423023389, 0423023394, 0423023399, 0423023451, 0423422042, 0423422045, 0423422048, 0423924004, 0424224127, 0424224131, 0424323096, 0426223004, 0426223008, 0427922038, 0427922042, 0428824102, 0428924075, 0428924079, 0428924083, 0429823019, 0429823023, 0430323199, 0430523194, 0430523197, 0430722034, 0431123070, 0432324031, 0432324156, 0432324159, 0432324162, 0432324165, 0432722064, 0433724005, 0433724010, 0433724015, 0434023084, 0434023087, 0434023090, 0434023111, 0434023116, 0434023121, 0434622027, 0434622031, 0434622035, 0434722013, 0434722159, 0434722163, 0436223068, 0436624179, 0436624184, 0436624189, 0436624194;
Why it was recalled
The DFPM yoke assembly ("yoke assembly") may detach from the spring arm. Should the yoke assembly detach from the spring arm, the yoke assembly would not be expected to fully detach but rather hang from the wiring within the yoke assembly.
Root cause (FDA determination)
Process control
Action the firm took
Steris issued an URGENT MEDICAL DEVICE RECALL NOTICE dated 06/17/2025 to its consignees on 06/23/2025 disseminated via FedEx. The notice explained the issue, potential risk, and requested the following: User Action: Please ensure the following steps are completed: 1. Review the revised Preventive Maintenance Manual and Interval Based Check List language depending on which HarmonyAIR Surgical Lighting System model(s) are in use at your facility (A-Series and/or E-Series) (please reference Attachments B-E of the letter for maintenance instructions). 2. Immediately complete the maintenance activities as instructed, and then again annually going forward. 3. Complete the Medical Device Recall Response Form included with this letter. 4. Return the completed Response Form via email to: Regulatory_Compliance@STERIS.com, or via fax to 440-392-8963. 5. Distributors: If you have further distributed this product, please identify your Customer(s), and notify them of this recall.
Recalling firm
- Firm
- Steris Corporation
- Address
- 5960 Heisley Rd, Mentor, Ohio 44060-1834
Distribution
- Distribution pattern
- Worldwide - US Nationwide and Puerto Rico and the country of Canada.
Timeline
- Recall initiated
- 2025-06-23
- Posted by FDA
- 2025-07-25
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #214408. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.