FDA Device Recalls

Recalls /

#214670

Product

High Pressure Tubing: REF: 502211002, HP9720E/C, HP9721E/C

FDA product code
FPKTubing, Fluid Delivery
Device class
Class 2
Medical specialty
General Hospital
510(k) numbers
K883718
Affected lot / code info
REF/UDI-DI/Lot(Expiration): 502211002/I3116954(8/16/2028); HP9720E/C/00884450859889/I3116059(12/10/2027), I3127321(12/30/2027), I3137220(1/28/2028); HP9721E/C/00884450859896/I3116060(12/10/2027)

Why it was recalled

High pressure tubing may relax over time causing it to partially separate from the connector resulting in a void that can trap air in the line. Use of the affected product may result in the trapped air being introduced into the patient.

Root cause (FDA determination)

Process control

Action the firm took

On 5/5/2025, recall notices were emailed to customers who were asked to do the following: 1) Identify affected devices and quarantine, discontinue use and distribution. Ensure that applicable personnel within your organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them and note the quantity distributed on the customer response form. 4. Return all affected devices to the firm. 5. Complete, and return the customer response form via email to response@merit.com If you have questions, contact the firm's customer service at RESPONSE@merit.com or via phone at +1 800 356 3748, Hours: 6 am to 6 pm MST, Mon-Fri

Recalling firm

Firm
Merit Medical Systems, Inc.
Address
1600 W Merit Pkwy South, Jordan, Utah 84095-2416

Distribution

Distribution pattern
US Nationwide distribution in the states of OH, CA, SD, VA, AZ, MT, TX, MI, FL, TN, PA, NJ, MN, ND, IA, LA, UT, KY, IL, OK, ME, IN, NY, NM, SC, AL, ID, CT, WI, GA, MA, NC, AR, WA.

Timeline

Recall initiated
2025-06-05
Posted by FDA
2025-08-12
Status