—#215711
Product
Exactech Equinoxe, Reverse Shoulder Humeral Liners, REF: 320-38-00,145-DEG PE 38MM HUM LINER +0/ 320-38-03,145-DEG PE 38MM HUM LINER +2.5/ 320-42-00,145-DEG PE 42MM HUM LINER +0/ 320-42-03,145-DEG PE 42MM HUM LINER +2.5/
- FDA product code
- PHX — Shoulder Prosthesis, Reverse Configuration
- Device class
- Class 2
- Medical specialty
- Orthopedic
- 510(k) numbers
- K063569
- Affected lot / code info
- All Serial Numbers, REF: 320-38-00/UDI:10885862086655 REF: 320-38-03/UDI: 10885862086662 REF: 320-42-00/UDI: 10885862086693 REF: 320-42-03/UDI: 10885862086709
Why it was recalled
Reverse Shoulder humeral liners have an articular surface position outside of the dimensional specification within the labeling.
Root cause (FDA determination)
Process design
Action the firm took
On August 15, 2025, Exactech, Inc. initiated a "Urgent Voluntary Medical Device Recall". Exactech asked consignees to take the following actions: 1. Healthcare providers who have treated patients using the Exactech Equinoxe Humeral Liners subject to this recall should continue to follow those patients pursuant to the healthcare provider s standard of care. 2. Do Not Use / Quarantine any of these products in your inventory. 3. Segregate and return unused units via prepaid packaging to Exactech. 4. Complete the Recall Confirmation Form confirming product return or on-hand inventory. 5. Transmission of this Recall Notice: This notice must be passed on to all who need to be aware within your organization and any locations the affected products has been transferred to. This recall has been reported to the U.S. FDA and will be reported to other regulatory authorities as required.
Recalling firm
- Firm
- Exactech, Inc.
- Address
- 2320 Nw 66th Ct, Gainesville, Florida 32653-1630
Distribution
- Distribution pattern
- Worldwide - US Nationwide distribution in the states of CA, CO, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MN, NV, NY, OH, OK, PR, SC, TN, TX, VA, WI and the countries of Australia, Austria, Brazil, Canada, Colombia, France, United Kingdom, Guatemala, India, Italy, Japan, Korea, Germany, The Netherlands, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Dubai.
Timeline
- Recall initiated
- 2025-08-15
- Posted by FDA
- 2025-09-12
- Status
- —