FDA Device Recalls

Recalls /

#215713

Product

Exactech Equinoxe, Reverse Shoulder Humeral Liners, 322-38-00 (145-DEG PE 38MM HUM LINER +0), REF: 322-38-03 (145-DEG PE 38MM HUM LINER +2.5), REF:322-42-00 (145-DEG PE 42MM HUM LINER +0), REF: 322-42-03 (145-DEG PE 42MM HUM LINER +2.5)

FDA product code
PHXShoulder Prosthesis, Reverse Configuration
Device class
Class 2
Medical specialty
Orthopedic
510(k) numbers
K223833
Affected lot / code info
All Serial Numbers, REF: 322-38-00/UDI:10885862593832, REF: 322-38-03/UDI:10885862593849, REF: 322-42-00/UDI: 10885862593917, REF:322-42-03/UDI:10885862593924

Why it was recalled

Reverse Shoulder humeral liners have an articular surface position outside of the dimensional specification within the labeling.

Root cause (FDA determination)

Process design

Action the firm took

On August 15, 2025, Exactech, Inc. initiated a "Urgent Voluntary Medical Device Recall". Exactech asked consignees to take the following actions: 1. Healthcare providers who have treated patients using the Exactech Equinoxe Humeral Liners subject to this recall should continue to follow those patients pursuant to the healthcare provider s standard of care. 2. Do Not Use / Quarantine any of these products in your inventory. 3. Segregate and return unused units via prepaid packaging to Exactech. 4. Complete the Recall Confirmation Form confirming product return or on-hand inventory. 5. Transmission of this Recall Notice: This notice must be passed on to all who need to be aware within your organization and any locations the affected products has been transferred to. This recall has been reported to the U.S. FDA and will be reported to other regulatory authorities as required.

Recalling firm

Firm
Exactech, Inc.
Address
2320 Nw 66th Ct, Gainesville, Florida 32653-1630

Distribution

Distribution pattern
Worldwide - US Nationwide distribution in the states of CA, CO, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MN, NV, NY, OH, OK, PR, SC, TN, TX, VA, WI and the countries of Australia, Austria, Brazil, Canada, Colombia, France, United Kingdom, Guatemala, India, Italy, Japan, Korea, Germany, The Netherlands, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Dubai.

Timeline

Recall initiated
2025-08-15
Posted by FDA
2025-09-12
Status

Source: openFDA Device Recall endpoint. Recall record ID #215713. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.