FDA Device Recalls

Recalls /

#215959

Product

Juggerknot Mini Soft Anchors, Model/Catalog Number: 912080

FDA product code
MBIFastener, Fixation, Nondegradable, Soft Tissue
Device class
Class 2
Medical specialty
Orthopedic
510(k) numbers
K110879
Affected lot / code info
UDI-DI: 00880304520950; Lot Number 0002713620

Why it was recalled

The outer carton box label is for the JuggerKnot 1.0mm Mini products. However, the products within the carton are JuggerKnot 1.4mm Short devices.

Root cause (FDA determination)

Under Investigation by firm

Action the firm took

An URGENT MEDICAL DEVICE RECALL notification letter dated 9/9/25 was sent to customers. Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the recall notice and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. Distributor Responsibilities: 1. Review this notification and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Immediately return all affected product from your distributorship and from affected hospitals within your territory. a. Complete Attachment 1 Inventory Return Certification Form for each return and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form must be returned even if you do not have affected products available to return in your territory. b. Include a hardcopy of Attachment 1 in each carton of your return shipment for immediate processing. c. M

Recalling firm

Firm
Biomet, Inc.
Address
56 E Bell Dr, Warsaw, Indiana 46582-6989

Distribution

Distribution pattern
US Nationwide distribution.

Timeline

Recall initiated
2025-09-09
Posted by FDA
2025-09-23
Status

Source: openFDA Device Recall endpoint. Recall record ID #215959. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.