Recalls / —
—#216044
Product
Salem Sump PVC Tubes: Product Code/Product Name: 8888265108 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 10 Fr/Ch (3.3 mm), 36 (91 cm) 8888265116 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 12 Fr/Ch (4.0 mm), 48 (122 cm) 8888265124 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 14 Fr/Ch (4.7 mm), 48 (122 cm) 8888265132 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 16 Fr/Ch (5.3 mm), 48 (122 cm) 8888265140 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 18 Fr/Ch (6.0 mm), 48 (122 cm) 8888266213 Salem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 10 Fr/Ch (3.3 mm) x 36 (91 cm) 8888266221 Salem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 12 Fr/Ch (4.0 mm) x 48 (122 cm) 8888266247 Salem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 14 Fr/Ch (4.7 mm) x 48 (122 cm) 8888266262 Salem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 16 Fr/Ch (5.3 mm) x 48 (122 cm) 8888266288 Salem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 18 Fr/Ch (6.0 mm) x 48 (122 cm)
- FDA product code
- KNT — Tubes, Gastrointestinal (And Accessories)
- Device class
- Class 2
- Medical specialty
- Gastroenterology, Urology
- 510(k) numbers
- K830710
- Affected lot / code info
- Product Code/Product Name/UDI-DI: 8888265108 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 10 Fr/Ch (3.3 mm), 36 (91 cm) 10884521000766 8888265116 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 12 Fr/Ch (4.0 mm), 48 (122 cm) 10884521020078 8888265124 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 14 Fr/Ch (4.7 mm), 48 (122 cm) 10884521020085 8888265132 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 16 Fr/Ch (5.3 mm), 48 (122 cm) 10884521020092 8888265140 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 18 Fr/Ch (6.0 mm), 48 (122 cm) 10884521020108 8888266213 Salem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 10 Fr/Ch (3.3 mm) x 36 (91 cm) 10192253012781 8888266221 Salem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 12 Fr/Ch (4.0 mm) x 48 (122 cm) 10884521000803 8888266247 Salem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 14 Fr/Ch (4.7 mm) x 48 (122 cm) 10884521000827 8888266262 Salem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 16 Fr/Ch (5.3 mm) x 48 (122 cm) 10884521000841 8888266288 Salem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 18 Fr/Ch (6.0 mm) x 48 (122 cm) 10884521000858 All Lots
Why it was recalled
Firm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use. Expansion of RES 95214.
Root cause (FDA determination)
Use error
Action the firm took
On September 18, 2025, the firm notified customers via Urgent Medical Device Product Correction Update letters. This is an update to the notice sent in July 2024 for RES 95214. The present recall includes all lots of additional product codes listed. The letter provides customers with additional guidance on how to properly use the Anti-Reflux Valve (ARV): Important Instructions: After placement of the Salem Sump tube, always inject 10ml-20ml of air before seating the Anti Reflux Valve (ARV) into the blue sump vent lumen. Instructions: 1. Seat the BLUE END of the ARV into the blue sump vent lumen of the Salem Sump tube. Insert the ARV gently into the Salem Sump tube to avoid difficult removal. 2. To cap tube, seat the WHITE END of the ARV into the suction lumen. Insert the ARV gently into the Salem Sump tube to avoid difficult removal. 3. To remove, pull valve in the same direction of assembly. Do not pull at an angle to avoid breaking the valve. Please visit https://www.mycardinalmsds.com for updated electronic IFUs.
Recalling firm
- Firm
- Cardinal Health 200, LLC
- Address
- 3651 Birchwood Dr, Waukegan, Illinois 60085-8337
Distribution
- Distribution pattern
- Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR, GU and the countries of Australia, Austria, Bahrain, Belgium, Bulgaria, Canada, Chile, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, French Polynesia, Germany, Greece, Hong Kong, Hungary, Ireland, Israel, Italy, Jordan, Africa, Kuwait, Luxembourg, Mexico Monaco, Netherlands, New Zealand, Northern Ireland, Norway, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Taiwan, UAE, United Kingdom, Vietnam.
Timeline
- Recall initiated
- 2025-09-18
- Posted by FDA
- 2025-10-17
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #216044. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.