FDA Device Recalls

Recalls /

#216050

Product

Salem Sump PVC Tubes: Product Code/Product Name: 1180264408 Salem Sump" PVC Gastroduodenal Tube, 10 Fr/Ch (3.3 mm) x 90 cm 1180264416 Salem Sump" PVC Gastroduodenal Tube, 12 Fr/Ch (4.0 mm) x 114 cm 1180264424 Salem Sump" PVC Gastroduodenal Tube, 14 Fr/Ch (4.7 mm) x 114 cm 1180264432 Salem Sump" PVC Gastroduodenal Tube, 16 Fr/Ch (5.3 mm) x 114 cm 1180264440 Salem Sump" PVC Gastroduodenal Tube, 18 Fr/Ch (6.0 mm) x 114 cm 1180264457 Salem Sump" PVC Gastroduodenal Tube, 20 Fr/Ch (6.7 mm) x 114 cm 1180264523 Salem Sump" Polyurethane Gastroduodenal Tube, 14 Fr/Ch (4.7 mm) x 114 cm 1180264531 Salem Sump" Polyurethane Gastroduodenal Tube, 16 Fr/Ch (5.3 mm) x 114 cm 1180264549 Salem Sump" Polyurethane Gastroduodenal Tube, 18 Fr/Ch (6.0 mm) x 114 cm 1180264556 Salem Sump" Polyurethane Gastroduodenal Tube, 20 Fr/Ch (6.7 mm) x 114 cm 1180264895 Salem Sump PVC Gastroduodenal Tube, 20 Fr/Ch (3.3mm), 108 cm 8888264804 Salem Sump PVC Gastroduodenal Tube, 10Fr/Ch (3.3 mm) x 85 cm 8888264820 Salem Sump PVC Gastroduodenal Tube, 12 Fr/Ch (4.0 mm) x 108 cm 8888264846 Salem Sump PVC Gastroduodenal Tube, 14 Fr/Ch (4.7mm), 108 cm 8888264861 Salem Sump PVC Gastroduodenal Tube, 16 Fr/Ch (5.3mm), 108 cm 8888264887 Salem Sump PVC Gastroduodenal Tube, 18 Fr/Ch (6.0mm), 108 cm SKUs not distributed in US

FDA product code
FEGTube, Double Lumen For Intestinal Decompression And/Or Intubation
Device class
Class 2
Medical specialty
Gastroenterology, Urology
Affected lot / code info
Product Code/Product Name/UDI-DI: 1180264408 Salem Sump" PVC Gastroduodenal Tube, 10 Fr/Ch (3.3 mm) x 90 cm 10192253006940 1180264416 Salem Sump" PVC Gastroduodenal Tube, 12 Fr/Ch (4.0 mm) x 114 cm 10192253006957 1180264424 Salem Sump" PVC Gastroduodenal Tube, 14 Fr/Ch (4.7 mm) x 114 cm 10192253006964 1180264432 Salem Sump" PVC Gastroduodenal Tube, 16 Fr/Ch (5.3 mm) x 114 cm 10192253006971 1180264440 Salem Sump" PVC Gastroduodenal Tube, 18 Fr/Ch (6.0 mm) x 114 cm 10192253007053 1180264457 Salem Sump" PVC Gastroduodenal Tube, 20 Fr/Ch (6.7 mm) x 114 cm 10192253006568 1180264523 Salem Sump" Polyurethane Gastroduodenal Tube, 14 Fr/Ch (4.7 mm) x 114 cm 10192253007060 1180264531 Salem Sump" Polyurethane Gastroduodenal Tube, 16 Fr/Ch (5.3 mm) x 114 cm 10192253006988 1180264549 Salem Sump" Polyurethane Gastroduodenal Tube, 18 Fr/Ch (6.0 mm) x 114 cm 10192253006995 1180264556 Salem Sump" Polyurethane Gastroduodenal Tube, 20 Fr/Ch (6.7 mm) x 114 cm 10192253007008 1180264895 Salem Sump PVC Gastroduodenal Tube, 20 Fr/Ch (3.3mm), 108 cm 10192253007015 8888264804 Salem Sump PVC Gastroduodenal Tube, 10Fr/Ch (3.3 mm) x 85 cm 10192253012408 8888264820 Salem Sump PVC Gastroduodenal Tube, 12 Fr/Ch (4.0 mm) x 108 cm 10192253012415 8888264846 Salem Sump PVC Gastroduodenal Tube, 14 Fr/Ch (4.7mm), 108 cm 10192253012422 8888264861 Salem Sump PVC Gastroduodenal Tube, 16 Fr/Ch (5.3mm), 108 cm 10192253012439 8888264887 Salem Sump PVC Gastroduodenal Tube, 18 Fr/Ch (6.0mm), 108 cm 10192253012446 All Lots

Why it was recalled

Firm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use. Expansion of RES 95214.

Root cause (FDA determination)

Use error

Action the firm took

On September 18, 2025, the firm notified customers via Urgent Medical Device Product Correction Update letters. This is an update to the notice sent in July 2024 for RES 95214. The present recall includes all lots of additional product codes listed. The letter provides customers with additional guidance on how to properly use the Anti-Reflux Valve (ARV): Important Instructions: After placement of the Salem Sump tube, always inject 10ml-20ml of air before seating the Anti Reflux Valve (ARV) into the blue sump vent lumen. Instructions: 1. Seat the BLUE END of the ARV into the blue sump vent lumen of the Salem Sump tube. Insert the ARV gently into the Salem Sump tube to avoid difficult removal. 2. To cap tube, seat the WHITE END of the ARV into the suction lumen. Insert the ARV gently into the Salem Sump tube to avoid difficult removal. 3. To remove, pull valve in the same direction of assembly. Do not pull at an angle to avoid breaking the valve. Please visit https://www.mycardinalmsds.com for updated electronic IFUs.

Recalling firm

Firm
Cardinal Health 200, LLC
Address
3651 Birchwood Dr, Waukegan, Illinois 60085-8337

Distribution

Distribution pattern
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR, GU and the countries of Australia, Austria, Bahrain, Belgium, Bulgaria, Canada, Chile, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, French Polynesia, Germany, Greece, Hong Kong, Hungary, Ireland, Israel, Italy, Jordan, Africa, Kuwait, Luxembourg, Mexico Monaco, Netherlands, New Zealand, Northern Ireland, Norway, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Taiwan, UAE, United Kingdom, Vietnam.

Timeline

Recall initiated
2025-09-18
Posted by FDA
2025-10-17
Status