Recalls / —
—#218229
Product
10mL Non-Sterile BD Luer-Lok Tip Syringe Only. Catalog Number: 301029. 850 bulk non-sterile units per case.
- FDA product code
- FMF — Syringe, Piston
- Device class
- Class 2
- Medical specialty
- General Hospital
- 510(k) numbers
- K980987
- Affected lot / code info
- Catalog Number: 301029., UDI-DI: N/A. Lot Number: 5268614. Expiration Date: 08/31/2030.
Why it was recalled
During the bulk packaging process, some Luer Slip syringes were incorrectly included in a batch designated for Luer Lok syringes.
Root cause (FDA determination)
Under Investigation by firm
Action the firm took
Becton Dickinson notified consignees on about 01/27/2026 via emailed URGENT: Medical Device Recall (Correction) letter. Consignees were instructed to visually verify that the correct syringe is selected for the intended clinical application prior to use and that it forms a secure connection, destroy any luer slip syringes found in inventory, circulate the notification to all personnel who need to be aware or to any organization where the potentially affected products have been transferred, and complete and return the provided Customer Response Form. Distributors and kit manufactures were instructed to identify all customers within their distribution network that purchased any affected product and provide a copy of the customer letter to all customers to advise them of this field action notification on BD s behalf. Becton Dickinson is investigating the root cause and will take corrective actions to prevent recurrence of the issue as well as will provide replacement for all discarded inventory.
Recalling firm
- Firm
- Becton Dickinson & Company
- Address
- 1 Becton Dr, Franklin Lakes, New Jersey 07417-1815
Distribution
- Distribution pattern
- US Nationwide distribution in the states of California and Texas.
Timeline
- Recall initiated
- 2026-01-27
- Posted by FDA
- 2026-03-06
- Status
- —