2.4 VOLT TAP LOCKING SCREW TAP TO 90MM. Part Number: 03.424.124-US.

FDA product code
HWXTap, Bone
Device class
Class 1
Medical specialty
Orthopedic
Affected lot / code info
Part Number: 03.424.124-US. GTIN: 10886982335944. Lot Number: 68900P9.

Why it was recalled

A full lot swap between VOLT Locking Screw Tap 03.424.124 lot 68900P9 and Cortical Screw Tap 03.424.126 lot 68982P8, resulting in the bones taps having an incorrect thread.

Root cause (FDA determination)

Process change control

Action the firm took

Depuy Synthes initially notified Sales Consultants and Regional Managers on about 02/25/2026, followed by notification of consignees. Consignees were instructed to examine inventory for affected units, quarantine them immediately, and contact Depuy Synthes sales consultant or customer support services to arrange the return/credits for affected units. Additionally, consignees were requested to complete and return the provided Business Response Form as well as notify all applicable personnel within the facility or any consignees if the affected units were further distributed. Consignees were also requested to post the notification in a visible area for awareness.

Recalling firm

Firm
Synthes (USA) Products LLC
Address
1301 Goshen Pkwy, West Chester, Pennsylvania 19380-5986

Distribution

Distribution pattern
US Nationwide distribution in the states of AR, AZ, CA, CO, FL, IN, LA, MD, MN, NC, NJ, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI.

Timeline

Recall initiated
2026-02-25
Posted by FDA
2026-03-30
Status