Product

Baxter System 1000 Single Patient Hemodialysis Delivery System; Model SYS1000; all series 1000 instruments labeled as System 1000, AltraTouch 1000, Baxter Tina and Baxter Aurora; Baxter Healthcare Corporation, Deerfield, IL 60015 USA or Baxter Healthcare Corporation, McGaw Park, IL 60085 USA

FDA product code
FKPSystem, Dialysate Delivery, Single Patient
Device class
Class 2
Medical specialty
Gastroenterology, Urology
510(k) numbers
K910215
Affected lot / code info
Serial numbers 1001S to 1142S; 50001 to 52109; 01001 to 23764

Why it was recalled

The dialysis treatment may be run without the bicarbonate due to an undetected bicarbonate pump stall following a sodium change.

Root cause (FDA determination)

Other

Action the firm took

Baxter sent recall letters dated 3/20/03 to the Medical Director and Hemodialysis Chief Technicians via first class mail on 3/20/03. The letters informed the accounts of the potential of running a treatment without bicarbonate due to an undetected bicarbonate pump stall following a sodium change, and provided instructions to change the B-A-alarm limits on the System 1000 Delivery Systems and confirm proper functionality after completing these changes. These changes will prevent undetected acid-only delivery in the event of a sodium level change

Recalling firm

Firm
Baxter Healthcare Corp. Rt
Address
120 & Wilson Rd, Round Lake, Illinois 60073

Distribution

Distribution pattern
The devices were distributed nationwide and internationally to Baxter subsidiaries in Argentina, Belgium, Bolivia, Brazil, Canada, Chile, China, Ecuador, Egypt, El Salvador, Guatemala, Honduras, Hong Kong, India, Mexico, Philippines, Singapore, South Korea, Thailand, Turkey, United Kingdom, Venezuela and Vietnam.

Timeline

Recall initiated
2003-03-20
Posted by FDA
2003-05-15
Terminated
2004-01-23
Status