FDA Device Recalls

Recalls /

#27063

Product

Elecsys Troponin T; catalog number 2017644

FDA product code
MMIImmunoassay Method, Troponin Subunit
Device class
Class 2
Medical specialty
Clinical Chemistry
510(k) numbers
K984105
Affected lot / code info
All lots.

Why it was recalled

Hemoglobin interference limitation revised downward; hemolyzed samples may result in false negative results.

Root cause (FDA determination)

Other

Action the firm took

An Urgent Product Correction letter was mailed to each account, during the week of 4/21/03, advising them not to use hemolyzed specimens with a hemoglobin level equal to, or greater than, 0.1 g/dL.

Recalling firm

Firm
Roche Diagnostics Corp.
Address
9115 Hague Road, Indianapolis, Indiana 46250-0457

Distribution

Distribution pattern
United States

Timeline

Recall initiated
2003-04-25
Posted by FDA
2003-05-29
Terminated
2003-06-19
Status

Source: openFDA Device Recall endpoint. Recall record ID #27063. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.