Product
Clearlink System Y-Type Blood/Solution Set, 10 drops/mL, 112'', Standard Blood Filter, 170 to 260 Micron Filter, Luer Activated Valve, Male Luer Lock Adapter, catalog 2C8750; Baxter Healthcare Corporation, Deerfield, IL 60015 USA
- FDA product code
- KSB — Set, Transfer (Blood/Plasma)
- Device class
- Class 2
- Medical specialty
- Hematology
- 510(k) numbers
- K003225, K993120
- Affected lot / code info
- catalog 2C8750, lot numbers UR128876 through UR168104. Please note: Baxter lot numbers are represented by an internal reference code ''UR'' followed by a six digit sequential number, or by the letter ''U'' followed by a six digit sequential number with and ''R'' suffix. Any lot number appearing within the identified six digit sequential number range is subject to the recall.
Why it was recalled
There are possible pinholes or perforations in the set tubing, caused by damage that occurred during the packaging process.
Root cause (FDA determination)
Other
Action the firm took
Recalled by letter dated 7/16/03. The accounts were informed of the potential for leaks in the set tubing, and asked to inspect their inventory for the affected product codes and lots, segregate any affected products found and discard them according to local procedures. The customers will receive credit for the discarded affected product.
Recalling firm
- Firm
- Baxter Healthcare Corp. Rt
- Address
- 120 & Wilson Rd, Round Lake, Illinois 60073
Distribution
- Distribution pattern
- Nationwide, Mexico, Canada, Columbia, Australia, New Zealand and Taiwan
Timeline
- Recall initiated
- 2003-07-16
- Posted by FDA
- 2003-08-21
- Terminated
- 2004-06-29
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #28408. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.