Product
Vitros Chemistry *** GLU (glucose) DT Slides ***25 slides per box *** CAT 153 2316 *** Ortho-Clinical Diagnostics Inc. , 100 Indigo Creek Dr., Rochester, NY 14626. [NOTE: product has a 24-month shelf life.]
- FDA product code
- CGA — Glucose Oxidase, Glucose
- Device class
- Class 2
- Medical specialty
- Clinical Chemistry
- 510(k) numbers
- K841503
- Affected lot / code info
- All lots with an expiration date of 1June 2005 and earlier.
Why it was recalled
Results of glucose measurements on sodium fluoride/potassium oxalate specimens can be negatively biased when using Vitros GLU and GLU DT slides stored in the refrigerator.
Root cause (FDA determination)
Other
Action the firm took
GLU Slides: Letters dated between 7/28/03 and 8/11/03 provide revised storage instructions. Customers using NaF/KOx blood collection tubes and have VITROS GLU slides that have been refrigerated for more than four months will receive replacement product. Those customers are instructed to continue to use the slides only until replacement product arrives. GLU DT Slides: Letters dated between 7/28/03 and 8/11/03 with the same instructions as for GLU slides.
Recalling firm
- Firm
- Ortho-Clinical Diagnostics
- Address
- 100 Indigo Creek Drive, Rochester, New York 14650-0881
Distribution
- Distribution pattern
- Nationwide and worldwide distribution to distributors and end-users.
Timeline
- Recall initiated
- 2003-07-28
- Posted by FDA
- 2003-09-11
- Terminated
- 2003-11-24
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #29041. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.