FDA Device Recalls

Recalls /

#31731

Product

Cytomics MXP Software, Version 1.0, Part No. 623688.

FDA product code
GKLCounter, Cell, Automated (Particle Counter)
Device class
Class 2
Medical specialty
Hematology
Affected lot / code info
Version 1.0

Why it was recalled

Software anomaly. Sample ID and the Run Date may become fixed within a header on the FlowPAGE printout. All samples subsequently run, with results printed using the FlowPAGE print format, will have the fixed Sample ID and Run Date, rather than the correct Sample ID and Run Date.

Root cause (FDA determination)

Other

Action the firm took

Firm sent modified operating instructions to users on 13 February 2004.

Recalling firm

Firm
Beckman Coulter Inc
Address
200 S Kraemer Blvd, Brea, California 92821-6208

Distribution

Distribution pattern
Nationwide and Canada

Timeline

Recall initiated
2004-02-13
Posted by FDA
2004-07-20
Terminated
2012-05-08
Status

Source: openFDA Device Recall endpoint. Recall record ID #31731. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.