Recalls / —
—#33491
Product
Cook brand Zenith AAA Converter with the H&L-B One-Shot Introducer System, preloaded converter - proximal diameter 24 mm - distal diameter 12 mm - length 80 mm; Reorder no. ESC-24-12-80.
- FDA product code
- MIH — System, Endovascular Graft, Aortic Aneurysm Treatment
- Device class
- Class 3
- Medical specialty
- Unknown
- PMA numbers
- P020018
- Affected lot / code info
- Lots 1356524, 1356532, 1356519, 1356525, 1356523, 1356522 and 1356529.
Why it was recalled
Introducer tubing may contain a vein of unextruded material and carbon black or a strand of tubing material may be displaced from the introducer during use.
Root cause (FDA determination)
Other
Action the firm took
The firm issued recall letters to customers on 6/10/04. The recall was extended and URGENT EXTENSION OF PRODUCT RECALL notices were sent out on June 23, 2004.
Recalling firm
- Firm
- Cook, Inc.
- Address
- 750 Daniels Way, Bloomington, Indiana 47404-9120
Distribution
- Distribution pattern
- Nationwide, Canada, Denmark and Japan.
Timeline
- Recall initiated
- 2004-06-10
- Posted by FDA
- 2004-07-27
- Terminated
- 2004-11-01
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #33491. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.