Recalls / —
—#34173
Product
BBL Columbia C.N.A. Agar w/5% Sheep Blood/Levine EMB - Catalog Number 295618.
- FDA product code
- JSO — Culture Media, Antimicrobial Susceptibility Test, Excluding Mueller Hinton Agar
- Device class
- Class 2
- Medical specialty
- Microbiology
- Affected lot / code info
- Lot number: 4153593 and Expiration date: August 26, 2004
Why it was recalled
Sheep Blood products may be contaminated with a Brucella species (B. ovis).
Root cause (FDA determination)
Other
Action the firm took
Distributors were notified by fax, all customers were notified by fax with a UPS overnight delivery of hard copy of recall letter to same customers. BD International was notified be e-mail. Distributor fax indicated that BD required their customer lists immediately due to the nature of the recall event. BD received all Distributor customer lists on the same day (7/2/04). Will monitor the number of consignees non-responding to the recall through sending a letter that contains a form that we ask the end-user to return. Response forms will be reconciled. Distribution of the affected lots ceased on 7/1/04. Customers are asked to discard their inventory.
Recalling firm
- Firm
- Becton Dickinson & Co.
- Address
- 7 Loveton Cir, Sparks, Maryland 21152-9212
Distribution
- Distribution pattern
- 83 (76 end-users, 3 international BD facilities and 4 distributors) worldwide.
Timeline
- Recall initiated
- 2004-07-02
- Posted by FDA
- 2004-08-07
- Terminated
- 2004-12-21
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #34173. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.