FDA Device Recalls

Recalls /

#34884

Product

Siregraph CF X-Ray System. Fluoroscopic image intensified. Model number 4466041

FDA product code
JAASystem, X-Ray, Fluoroscopic, Image-Intensified
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K960266
Affected lot / code info
Serial number: 3131.

Why it was recalled

X-ray tube movement when table is tilted more than 90 degrees

Root cause (FDA determination)

Other

Action the firm took

The recalling firm issued a Customer DAfety Advisory letter dated 6/7/04 to the hospitals. The letter informs them of the potential problem and provides preventative measures to avoid its occurance.

Recalling firm

Firm
Siemens Medical Solutions USA, Inc
Address
51 Valley Stream Pkwy, Malvern, Pennsylvania 19355-1406

Distribution

Distribution pattern
The products were shipped to medical facilities in FL, IL, MI, NH, NJ, PA, and TX.

Timeline

Recall initiated
2004-06-07
Posted by FDA
2004-09-18
Terminated
2005-08-18
Status

Source: openFDA Device Recall endpoint. Recall record ID #34884. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.