Product
POLYSTAR T.O.P., Angiographic X-Ray System, Model Numbers 1148902 and 4784505.
- FDA product code
- IZI — System, X-Ray, Angiographic
- Device class
- Class 2
- Medical specialty
- Radiology
- 510(k) numbers
- K913120
- Affected lot / code info
- Serial Numbers for model 1148902: 01000 through 01267, 02001 through 02117. Serial numbers for model4784505: 03001 through 035158, 04001 through 04136, and 05001 through 05098
Why it was recalled
Orbital gears may exhibit excess wear
Root cause (FDA determination)
Other
Action the firm took
The recalling firm issued Update Instructions, AX077/04/S, to Siemen''s Customer Service Representative (CSR) to visit affected sites and check the orbital gears for excessive wear. If the CSR finds gears exhibiting excessive wear the orbital gears will be replaced. If no wear is observed the gears will be lubricated as a preventive measure.
Recalling firm
- Firm
- Siemens Medical Solutions USA, Inc
- Address
- 51 Valley Stream Pkwy, Malvern, Pennsylvania 19355-1406
Distribution
- Distribution pattern
- The product was shipped to medical facilities nationwide and government accounts in KS and MI.
Timeline
- Recall initiated
- 2004-09-23
- Posted by FDA
- 2004-11-16
- Terminated
- 2005-01-25