Product
Sabraset Administration Sets with AutoClamp; for use with model 6060 multi-therapy pumps only; a prescription, sterile, non-pyrogenic fluid pathway; Manufactured for Baxter Healthcare Corporation, Deerfield, IL 60015 U.S.A., the following sets were made in Mexico: a) Sabraset 56050 100 Administration Set with 100 mL Bag and Cassette, 56'', Volume 3 mL, 100 mL Bag with Fill Port, Cassette with AutoClamp Device, Slide Clamp, Male Luer Lock Adapter b) Sabraset 56050 250 Administration Set with 2500 mL Bag and Cassette, 53'', Volume 3 mL, 250 mL Bag with Fill Port, Cassette with AutoClamp Device, Slide Clamp, Male Luer Lock Adapter
- FDA product code
- FPA — Set, Administration, Intravascular
- Device class
- Class 2
- Medical specialty
- General Hospital
- 510(k) numbers
- K002679
- Affected lot / code info
- Product codes 560500-100 and 560500-250, all lot numbers other than lots produced within the range of 417519 to 499999, or those with an adhesive circular colored sticker (Avery dot) at the top of the individual package on the Tyvek side or on the shipper label. Any lot numbers from the following obsolete Sabraset product codes are also being recalled: 5600PM, 560100, 560101, 560110, 560111, 560112, 560112-GEL, 560112-L, 560113, 560115, 560116, 560122-L, 561101, 567100, 567100-L, 567112-L, 567122-L, 569100, 569111, 569122, 560200-100, 560522-100
Why it was recalled
A missing platen assembly in the cassette of the solution sets results in the infusion pump being unable to occlude the tubing of the set. This can lead to free-flow of the solution and a possible overdose of medication to the patient.
Root cause (FDA determination)
Other
Action the firm took
Urgent Product Recall letters dated 11/19/04 were sent to the direct accounts on the same date via UPS ground. The accounts were informed that free-flow situations were associated with the listed product codes and lot numbers due to a missing platen assembly in the cassette of the solution sets, which could result in an overdose of medication to the patient. The accounts were requested to stop use of the affected lot numbers and product codes and return the sets to Baxter c/o NNC Group, 2670 Executive Dr., Indianapolis, IN 46241, following the enclosed instructions. Any questions concerning the return process were directed to NCC at 1-866-300-5085. Any medical or technical questions were directed to Baxter at 1-800-422-9837.
Recalling firm
- Firm
- Baxter Healthcare Corp. Rt.
- Address
- 120 & Wilson Rd, Round Lake, Illinois 60073
Distribution
- Distribution pattern
- Nationwide, including Puerto Rico, and internationally to Australia, Belgium, Canada, Hong Kong, Israel, Japan, Singapore, Sweden and Switzerland
Timeline
- Recall initiated
- 2004-11-19
- Posted by FDA
- 2005-01-08
- Terminated
- 2006-06-28
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #36055. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.