Recalls / —
—#36489
Product
Cook brand Zenith Flex AAA Endovascular graft with the H&L-B One-Shot Introduction System( A= 32mm, B=112mm, C=82mm, D=12mm, E= 20.0f, F=.035'', G =40cm); Reorder number TFBB-32-82.
- FDA product code
- MIH — System, Endovascular Graft, Aortic Aneurysm Treatment
- Device class
- Class 3
- Medical specialty
- Unknown
- PMA numbers
- P020018
- Affected lot / code info
- All units distributed on or before 12/17/04.
Why it was recalled
The firm has received complaints of difficulty in releasing the barbed suprarenal stent from the top cap that constrains it, which may result in misplacement of the graft.
Root cause (FDA determination)
Other
Action the firm took
Recall letter dated 12/20/04 was issued to each consignee via FedEx instructing them to cease using the products and to return them to the firm.
Recalling firm
- Firm
- Cook, Inc.
- Address
- 750 Daniels Way, Bloomington, Indiana 47404-9120
Distribution
- Distribution pattern
- Nationwide and Canada.
Timeline
- Recall initiated
- 2004-12-20
- Posted by FDA
- 2005-02-03
- Terminated
- 2005-12-02
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #36489. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.