Recalls / —
—#38306
Product
BD 25G Spinal Needles. (Spinal Anesthesia Needle)
- FDA product code
- BSP — Needle, Conduction, Anesthetic (W/Wo Introducer)
- Device class
- Class 2
- Medical specialty
- Anesthesiology
- Affected lot / code info
- Reorder Number, Lot Number: 400440, 4275101; 400440, 4282078; 400440, 4317744; 400479, 3196624; 400479, 4161815; 400479, 4254414; 400479, 4324795; 400496, 4173091; 400590, 3191978; 400590, 4121350; 400621, 4146482; 401586, 3182399; 405078, 3196618; 405078, 4169804; 405078, 4194109; 405078, 4268754; 405078, 4282079; 405078, 4324791; 405078, 4349021; 405138, 4173087; 405138, 4181398; 405140, 3224096; 405140, 4110295; 405140, 4237191; 405140, 4295872, 405140, 4314481; 405140, 4336365; 405170, 4201356; 405170, 4201357 405180, 4096782; 405180, 4145523; 405180, 4148675; 405234, 3174061; 405234, 4062807; 405234, 4105111; 405234, 4163586; 405234, 4265265; 405234, 4282715; 405234, 4282716; 405234, 4314480; 405234, 4358850; 409442, 3174062; 409442, 3226365; 409442, 4110275; 409442, 4121348; 409442, 4230810; 409442, 4230813; 409442, 4257418; 409442, 4300308; 409442, 4320423; 409442, 4329215; 409442, 4349941.
Why it was recalled
A significant level of 25G spinal needle hub leakage complaints was observed. Subsequent investigation found the leakage to be attributable to improper formation of the spinal needle hub, specifically an undersized taper and the prescence of a slight sink in the taper surface.
Root cause (FDA determination)
Other
Action the firm took
The firm sent out notification letters on 3/16/2005.
Recalling firm
- Firm
- Becton Dickinson & Company
- Address
- 1 Becton Dr, Franklin Lakes, New Jersey 07417-1815
Distribution
- Distribution pattern
- The firm uses distributors and ships the product directly to over 4000 hospitals and clinics nationwide.
Timeline
- Recall initiated
- 2005-03-16
- Posted by FDA
- 2005-06-25
- Terminated
- 2005-11-07
- Status
- —