Product
LifeShield Latex-Free HEMA Blood PlumSet, Nonvented, 110 Inch with 210 Micron Filter, Prepierced Injection Site and OPTION-LOK, Dual Channel, Capped Secondary Port; a sterile Rx blood administration set for use with the Plum Series Infuser; Hospira, Inc., Lake Forest, IL 60045 USA, Made in Costa Rica; list number 11235-03
- FDA product code
- BRZ — Set, Blood Transfusion
- Device class
- Class 2
- Medical specialty
- General Hospital
- 510(k) numbers
- K830014, K920736, K941214
- Affected lot / code info
- lot 25-086-5H
Why it was recalled
Inlet and outlet tubing on the cassette may be reversed, which would result in reverse flow.
Root cause (FDA determination)
Other
Action the firm took
Hospira sent product recall letters dated 5/6/05 to the accounts via Federal Express overnight letter on the same date. The letters informed the accounts of the possiblitiy of the inlet and outlet tubing being reversed on the cassette, which would allow the patient''s blood to be drawn up the tubing. The accounts were requested to examine their inventory for the affected lot, quarantine any found and call Hospira Customer Care at 1-877-946-7747 to arrange for return of the product for replacement.
Recalling firm
- Firm
- Hospira Inc.
- Address
- 275 N Field Dr, Lake Forest, Illinois 60045-2579
Distribution
- Distribution pattern
- Nationwide including Puerto Rico, and internationally to Mexico and Australia
Timeline
- Recall initiated
- 2005-05-06
- Posted by FDA
- 2005-11-24
- Terminated
- 2005-11-29
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #38878. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.