FDA Device Recalls

Recalls /

#39545

Product

Depuy brand LCS Complete metal backed patella, porocoat, LRG+; REF 1294-09-670.

FDA product code
NJLProsthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing
Device class
Class 3
Medical specialty
Unknown
PMA numbers
P830055/R44
Affected lot / code info
Lot 1851632.

Why it was recalled

There may be dissociation of the polyethylene from the baseplate because the inside diameter of the polyethylene articular bearing was manufactured to an oversized condition.

Root cause (FDA determination)

Other

Action the firm took

Hospitals were notified via letter dated June 10, 2005 requesting that they quarantine product on hand and send in a response card to the firm.

Recalling firm

Firm
Depuy Orthopaedics, Inc.
Address
700 Orthopaedic Dr, Warsaw, Indiana 46582-3994

Distribution

Distribution pattern
Nationwide and international.

Timeline

Recall initiated
2005-06-07
Posted by FDA
2005-07-21
Terminated
2006-05-02
Status

Source: openFDA Device Recall endpoint. Recall record ID #39545. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.

— recall: Depuy brand LCS Complete metal backed patella, porocoat, LRG+; REF 1294-09-670. · FDA Device Recalls