Product
Depuy P.F.C. E Knee System, stabilized all-polyethylene tibia, Sz 3, 10.0 mm; Ref 96-2641.
- FDA product code
- NJL — Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing
- Device class
- Class 3
- Medical specialty
- Unknown
- PMA numbers
- P830055S084
- Affected lot / code info
- Lots 83414A, 84168A, 84488A, 84665A, 85744A, 86011A, 86230A, 86325A, 86479A and 86704A.
Why it was recalled
An incomplete seal on the inner pouch allows exposure to oxygen, which may result in oxidation and would affect long-term performance of the implant.
Root cause (FDA determination)
Other
Recalling firm
- Firm
- Depuy Orthopaedics, Inc.
- Address
- 700 Orthopaedic Dr, Warsaw, Indiana 46582-3994
Distribution
- Distribution pattern
- Nationwide, Argentina, Brazil, Canada, Caguas, Colombia, Ireland, Malaysia, Mexico, Panama, Peru, Singapore, South Africa, Sweden, Uruguay and Venezuela.
Timeline
- Recall initiated
- 2005-05-26
- Posted by FDA
- 2005-08-12
- Terminated
- 2005-12-02