FDA Device Recalls

Recalls /

#42885

Product

Baxter 6060 Multi-Therapy Infusion Pump, product codes 2M9832, 2M9832P and 2M9832R, and Sabratek 6060 Homerun Infusion Pumps, product codes 606000, 606000-40, 606000-40L, and 606000-40I; Manufactured by an affiliate of Baxter Healthcare Corp., Deerfield, IL 60013 USA, made in Singapore

FDA product code
FRNPump, Infusion
Device class
Class 2
Medical specialty
General Hospital
510(k) numbers
K002679
Affected lot / code info
product codes 2M9832, 2M9832P and 2M9832R; product codes 606000, 606000-40, 606000-40L, and 606000-40I; 2M9832: serial numbers 501001HR-501263HR, 502001HR-502156HR, 503001HR-503155HR, 505001HR-505210HR, 507001HR-507054HR, 502001HG-502060HG, 503001HG-503048HG, 505001HG-505276HG, 506001HG-506180HG, 501001HJ-501072HJ, 502001HJ-502072HJ, 503001HJ-503060HJ, 504001HJ-504060HJ, 505001HJ-505060HJ, 501001HW-501020HW, 502001HW-502024HW, 501001HF-501012HF, 502001HF-502024HF

Why it was recalled

Possible shorting of some printed circuit boards could result in a possible interruption of therapy or cause the pump to fail the Power-On Self Test.

Root cause (FDA determination)

Other

Action the firm took

Baxter sent Urgent Device Correction letters dated 11/7/05 to the user accounts, advising them of the potential for the circuit boards in the 6060 pump to fail due to a problem that occurred in the supplier's manufacturing process. The letter included a list of 6060 pump serial numbers sent to the user that may contain an affected circuit board. To mitigate safety risks to the patient, Baxter advised customers to immediately discontinue use of the pump for therapies where interruption of the infusion could cause immediate patient harm. In addition, the pump is not to be used if it fails the Power-On Self Test. The accounts were instructed to contact Baxter Medication Delivery Services at 1-800-843-7867 for information regarding the return process for the affected 6060 pumps for correction. Baxter sent follow-up letters dated 2/16/06 to the accounts, to the attention of the Director of Biomedical Engineering, informing them that Baxter has sufficient quantities of replacement circuit boards to begin the repairs. The accounts were requested to review the enclosed list of serial numbers for accuracy and telephone Baxter at 1-800-843-7867 to schedule the return of any 6060 Infusion Pump that requires replacement of the circuit board.

Recalling firm

Firm
Baxter Healthcare Corp. Rt.
Address
120 & Wilson Rd, Round Lake, Illinois 60073

Distribution

Distribution pattern
Nationwide, and internationally in France, Germany, Sweden and teh United Kingdom.

Timeline

Recall initiated
2005-11-07
Posted by FDA
2006-01-28
Terminated
2007-06-19
Status

Source: openFDA Device Recall endpoint. Recall record ID #42885. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.