Recalls / —
—#43401
Product
Roche Elecsys total PSA CalSet, catalog number 11731696322.
- FDA product code
- LTJ — Prostate-Specific Antigen (Psa) For Management Of Prostate Cancers
- Device class
- Class 2
- Medical specialty
- Immunology
- PMA numbers
- P990056S001
- Affected lot / code info
- Lot 171735.
Why it was recalled
There is the possibility of high results with sera expected to have non-detectable levels of tPSA if these reagent lots are used in combination on the Modular Analytics E170 analyzer.
Root cause (FDA determination)
Other
Action the firm took
Consignees were notified of the incompatibility via reagent bulletin dated 12/7/05.
Recalling firm
- Firm
- Roche Diagnostics Corp.
- Address
- 9115 Hague Rd, Indianapolis, Indiana 46256-1025
Distribution
- Distribution pattern
- Nationwide.
Timeline
- Recall initiated
- 2005-12-07
- Posted by FDA
- 2006-01-11
- Terminated
- 2006-02-26
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #43401. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.