Product
Paradigm Link -- Blood Glucose Monitor.
- FDA product code
- NBW — System, Test, Blood Glucose, Over The Counter
- Device class
- Class 2
- Medical specialty
- Clinical Chemistry
- 510(k) numbers
- K041478
- Affected lot / code info
- All lots under the following Catalog Numbers: 322200; 322201; 322202; 322203; 322204; 322205; 322206; 322207.
Why it was recalled
BD Blood Glucose Monitors may inadvertently switch the Unit of Measure from mg/dL to mmol/L or vice versa during battery insertion or when a meter is dropped.
Root cause (FDA determination)
Other
Action the firm took
Recall notifications letters began mailing on 1/5/2006 to distributors, retailers, and DME providers. Healthcare professionals and consumers mailings will be during the week of 1/9/2006
Recalling firm
- Firm
- Becton Dickinson & Company
- Address
- 1 Becton Dr, Franklin Lakes, New Jersey 07417-1815
Distribution
- Distribution pattern
- The firm distributes their blood glucose monitors Nationwide to distributors, retailers, durable Medical Equipment providers, Healthcare Professionals, and consumers. Foreign accounts include locations in Canada and Germany. There is one VA account.