FDA Device Recalls

Recalls /

#44877

Product

SOMATOM CT System Emotion 16. Computed tomography x-ray system.

FDA product code
JAKSystem, X-Ray, Tomography, Computed
Device class
Class 2
Medical specialty
Radiology
510(k) numbers
K050297
Affected lot / code info
Emotion 16 model number 7734713 with serial numbers 39049, 39050, 39012, 39015 and 39077

Why it was recalled

When using the LungCare feature in the SOMATOM CT Systems some marks made on CT images may not be saved.

Root cause (FDA determination)

Other

Action the firm took

The recalling firm has issued a software patch to the affected customers per Update Instructions CT086/05/S and CT087/05/S. By letter 3/9/2006. In addition, a Siemens Service Engineer will visit affected sites to install this software patch which will resolve this issue.

Recalling firm

Firm
Siemens Medical Solutions USA, Inc
Address
51 Valley Stream Pkwy, Malvern, Pennsylvania 19355-1406

Distribution

Distribution pattern
Nationwide. The product was shipped to AK, AZ, CA, DE, FL, GA, IL, IN, KY, LA, MD, ME, MI, MT, NC, NY, OH, OK, PA, PR, TX, UT, VA, WA and WI.

Timeline

Recall initiated
2006-03-09
Posted by FDA
2006-04-26
Terminated
2006-07-19
Status

Source: openFDA Device Recall endpoint. Recall record ID #44877. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.

— recall: SOMATOM CT System Emotion 16. Computed tomography x-ray system. · FDA Device Recalls