Recalls / —
—#46884
Product
Roche Tina-Quant ASLO (Anti-streptolysin O) Reagent Kit; Catalog No. 1931601 (Roche Material No. 11931601216) Lot No. 674653
- FDA product code
- GTQ — Antistreptolysin - Titer/Streptolysin O Reagent
- Device class
- Class 1
- Medical specialty
- Microbiology
- Affected lot / code info
- Lot 674653; exp. 12/31/2006.
Why it was recalled
The vial containing the R2 reagent is labeled as R1 reagent and if placed in the R1 slot of the analyzer, erroneous results will be obtained. (extremely low patient and control results)
Root cause (FDA determination)
Other
Action the firm took
Consignees were notified via recall letter dated 7/25/06 to cease using and to discard the product.
Recalling firm
- Firm
- Roche Diagnostics Corp.
- Address
- 9115 Hague Rd, Indianapolis, Indiana 46256-1025
Distribution
- Distribution pattern
- Nationwide including the states of Arizona, California, New Jersey and Pennsylvania.
Timeline
- Recall initiated
- 2006-07-25
- Posted by FDA
- 2006-08-09
- Terminated
- 2007-04-18
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #46884. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.