FDA Device Recalls

Recalls /

#46884

Product

Roche Tina-Quant ASLO (Anti-streptolysin O) Reagent Kit; Catalog No. 1931601 (Roche Material No. 11931601216) Lot No. 674653

FDA product code
GTQAntistreptolysin - Titer/Streptolysin O Reagent
Device class
Class 1
Medical specialty
Microbiology
Affected lot / code info
Lot 674653; exp. 12/31/2006.

Why it was recalled

The vial containing the R2 reagent is labeled as R1 reagent and if placed in the R1 slot of the analyzer, erroneous results will be obtained. (extremely low patient and control results)

Root cause (FDA determination)

Other

Action the firm took

Consignees were notified via recall letter dated 7/25/06 to cease using and to discard the product.

Recalling firm

Firm
Roche Diagnostics Corp.
Address
9115 Hague Rd, Indianapolis, Indiana 46256-1025

Distribution

Distribution pattern
Nationwide including the states of Arizona, California, New Jersey and Pennsylvania.

Timeline

Recall initiated
2006-07-25
Posted by FDA
2006-08-09
Terminated
2007-04-18
Status

Source: openFDA Device Recall endpoint. Recall record ID #46884. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.