Product

Terumo Advanced Perfusion System 1; 220V -240V; Model 801764 (sold outside of the U.S. only).

FDA product code
DTQConsole, Heart-Lung Machine, Cardiopulmonary Bypass
Device class
Class 2
Medical specialty
Cardiovascular
Affected lot / code info
All units Serial numbers 0006 through 0213.

Why it was recalled

Improperly adjusting the occlusion setting may result in a pump jam error and pump stoppage.

Root cause (FDA determination)

Other

Action the firm took

A letter dated 11/10/06 was sent to consignees providing them with an addendum (814576R/B) to the operators'' manual. Users of systems produced prior to May 2006 received only this addendum, but users produced between May and November 2006 had previously received addendum 814576R/A which is corrected and to be replaced by the new addendum 814576R/B.

Recalling firm

Firm
Terumo Cardiovascular Systems Corp
Address
6200 Jackson Rd, Ann Arbor, Michigan 48103-9586

Distribution

Distribution pattern
Worldwide, including USA, Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand and United Arab Emirates.

Timeline

Recall initiated
2006-11-10
Posted by FDA
2006-12-27
Terminated
2008-06-19