Recalls / —
—#48572
Product
Terumo Advanced Perfusion System 1; 100V -120V, 15 A (circuit breaker), 50/60 Hz (20A power source required); Models 801763.
- FDA product code
- DTQ — Console, Heart-Lung Machine, Cardiopulmonary Bypass
- Device class
- Class 2
- Medical specialty
- Cardiovascular
- 510(k) numbers
- K022947
- Affected lot / code info
- Serial numbers 0093 and 0104 through 0114.
Why it was recalled
The gas system may fail calibration prior to use due to incorrect gas system flowmeter software.
Root cause (FDA determination)
Process control
Action the firm took
U.S. consignees were visited and corrections made between May 2004 and November 2005. Foreign affiliates were instructed to make corrections on international units, but the firm has not received confirmation yet that these units have been corrected.
Recalling firm
- Firm
- Terumo Cardiovascular Systems Corp
- Address
- 6200 Jackson Rd, Ann Arbor, Michigan 48103-9586
Distribution
- Distribution pattern
- Worldwide distribution, including USA, Belgium, Germany, Japan, Korea and Republic of China.
Timeline
- Recall initiated
- 2004-05-01
- Posted by FDA
- 2006-12-27
- Terminated
- 2009-07-14
- Status
- —