Product

Terumo Advanced Perfusion System 1 Air Bubble Detection Module; Catalog number 802110.

FDA product code
DTQConsole, Heart-Lung Machine, Cardiopulmonary Bypass
Device class
Class 2
Medical specialty
Cardiovascular
510(k) numbers
K022947
Affected lot / code info
Serial numbers 00023 through 00026, 00034, 00041, 00051, 00060, 00063 through 00065, 00067 through 00069, 00071 through 00089, 00091 through 00139, 00141 through 00242, 00246 through 00280, 00282 through 00297, 00299 through 00323, 00325 through 00410 and 00419.

Why it was recalled

Device module(s) may fail to properly initialize during start up, thus resulting in that single module being inoperable and a red 'X' being displayed on the monitor screen.

Root cause (FDA determination)

Other

Action the firm took

U.S. consignees were sent an addendum to the operators manual on October 12, 2004 instructing them to cycle power to the unit to clear the error. International distributors were requested to provide this information to their customers on 12/15/04, but Terumo has not yet received confirmation that this was completed.

Recalling firm

Firm
Terumo Cardiovascular Systems Corp
Address
6200 Jackson Rd, Ann Arbor, Michigan 48103-9586

Distribution

Distribution pattern
Worldwide Distribution-USA and countries Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand and United Arab Emirates.

Timeline

Recall initiated
2004-10-12
Posted by FDA
2006-12-30
Terminated
2007-02-01