Product
Terumo Advanced Perfusion System 1 Data Interface CDI 100; Catalog number 802558.
- FDA product code
- DTQ — Console, Heart-Lung Machine, Cardiopulmonary Bypass
- Device class
- Class 2
- Medical specialty
- Cardiovascular
- 510(k) numbers
- K022947
- Affected lot / code info
- Serial numbers 00008 through 00011, 00013, 00020 through 00030, 00032 through 00034, 00036 through 00059 and 00061 through 00078.
Why it was recalled
Device module(s) may fail to properly initialize during start up, thus resulting in that single module being inoperable and a red 'X' being displayed on the monitor screen.
Root cause (FDA determination)
Other
Action the firm took
U.S. consignees were sent an addendum to the operators manual on October 12, 2004 instructing them to cycle power to the unit to clear the error. International distributors were requested to provide this information to their customers on 12/15/04, but Terumo has not yet received confirmation that this was completed.
Recalling firm
- Firm
- Terumo Cardiovascular Systems Corp
- Address
- 6200 Jackson Rd, Ann Arbor, Michigan 48103-9586
Distribution
- Distribution pattern
- Worldwide Distribution-USA and countries Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand and United Arab Emirates.
Timeline
- Recall initiated
- 2004-10-12
- Posted by FDA
- 2006-12-30
- Terminated
- 2007-02-01
- Status
- —