Product
Terumo Advanced Perfusion System 1 Central Control Monitor; Catalog number 802100.
- FDA product code
- DTQ — Console, Heart-Lung Machine, Cardiopulmonary Bypass
- Device class
- Class 2
- Medical specialty
- Cardiovascular
- 510(k) numbers
- K022947
- Affected lot / code info
- Serial numbers 0047, 0049, 0052, 0054, 0055, 0065, 0066, 0069, 0071, 0072, 0074, 0076 through 0088, 0090 through 0111, 0113 through 0131, 0133 through 0276, 0278 through 0288, 0291 through 0293, 0295 through 0301 and 0303 through 0350.
Why it was recalled
Device module(s) may fail to properly initialize during start up, thus resulting in that single module being inoperable and a red 'X' being displayed on the monitor screen.
Root cause (FDA determination)
Other
Action the firm took
U.S. consignees were sent an addendum to the operators manual on October 12, 2004 instructing them to cycle power to the unit to clear the error. International distributors were requested to provide this information to their customers on 12/15/04, but Terumo has not yet received confirmation that this was completed.
Recalling firm
- Firm
- Terumo Cardiovascular Systems Corp
- Address
- 6200 Jackson Rd, Ann Arbor, Michigan 48103-9586
Distribution
- Distribution pattern
- Worldwide Distribution-USA and countries Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand and United Arab Emirates.
Timeline
- Recall initiated
- 2004-10-12
- Posted by FDA
- 2006-12-30
- Terminated
- 2007-02-01
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #48590. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.