Recalls / —
—#48745
Product
Terumo Advanced Perfusion System 1; Integrated (Sarns) Centrifugal System Control Unit; Catalog number 801046.
- FDA product code
- DTQ — Console, Heart-Lung Machine, Cardiopulmonary Bypass
- Device class
- Class 2
- Medical specialty
- Cardiovascular
- 510(k) numbers
- K022947
- Affected lot / code info
- Serial numbers 0020 through 0025, 0030 through 0036, 0038 through 0040, 0042, 0044, 0045, 0047 through 0102, 0106 through 0130, 0132 through 0155, 0157 through 0166, 0169 through 0216, 0218 through 0335 and 0338 through 0380.
Why it was recalled
The pump may lose communication with the central control monitor, causing the monitor to display a service message "System Computer Needs Service" and to become non-operational.
Root cause (FDA determination)
Other
Action the firm took
Consignees were notified of the problem via letter dated 11/10/06 and informed that they would be contacted to schedule a service call to install software upgrades. Installation of software upgrades began at customer locations on 10/16/06.
Recalling firm
- Firm
- Terumo Cardiovascular Systems Corp
- Address
- 6200 Jackson Rd, Ann Arbor, Michigan 48103-9586
Distribution
- Distribution pattern
- Worldwide distribution --- including USA and countries of Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand and United Arab Emirates.
Timeline
- Recall initiated
- 2006-10-16
- Posted by FDA
- 2007-02-14
- Terminated
- 2009-07-07
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #48745. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.