Recalls / —
—#48751
Product
M1020B PulseOximetry Module Philips FAST Sp02 and Pulse Oximetry Nellcor OxiMax Compatible used with the IntelliVue or M3/M4 patient monitor M1002
- FDA product code
- DQA — Oximeter
- Device class
- Class 2
- Medical specialty
- Cardiovascular
- 510(k) numbers
- K033715
- Affected lot / code info
- M1020B #AO01 AND #A02 with S/N DE524xxxxx M102B #A01 and #A02 with S/N DE612xxxxx With Sp02 Firmware (Fw) Rev: Sp02 DSP Fw-Rev. A.01.42
Why it was recalled
Unexpected pulse oximetry (Sp02) readings (100%) over time when sensor is not attached to patient. In addition, with a sensor attached to a patient, an incorrect high Sp02 may be displayed when a pulse rate is in a range of 185BPM
Root cause (FDA determination)
Other
Action the firm took
Philips issued a Customer Letter and Urgent Device Correction letter to consignees on 09/21/06. Users were notified of the problem and provided with Instructions to Mitigate Risk. A software upgrade to: version D.00.61 and Sp02 FW A.01.46 is mandatory.
Recalling firm
- Firm
- Philips Medical Systems
- Address
- 3000 Minuteman Road, Andover, Massachusetts 01810-1032
Distribution
- Distribution pattern
- Worldwide, including USA, Canada, Australia, China, Hong Kong, India, Indonesia, Japan, Phillipines, Singapore, Taiwan, Thailand, Vietnam, Argentina, Brazil, Chile, Colombia, El Salvador, Jamaica, Mexico, Nicaragua. Panama, Paraguay, Peru, Venezuela, Austria, Belgium, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Guadeloupe, Hungary, Germany, Israel, Italy, Jordan, Kenya, Kuwait, Lebanon, Lithuania, Luxembourg, Macedonia, Mauritania, Mauritius, Netherlands, Norway, Oman, Slovenia, South Africa, Spain, Switzerland, Turkey, UAE, UK, Yemen, Zambia, Zimbabwe
Timeline
- Recall initiated
- 2006-09-21
- Posted by FDA
- 2006-11-16
- Terminated
- 2020-08-11
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #48751. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.