Recalls / —
—#48924
Product
Allura X-per FD 10/10 x-ray, and fluoroscopy
- FDA product code
- IZI — System, X-Ray, Angiographic
- Device class
- Class 2
- Medical specialty
- Radiology
- 510(k) numbers
- K031333
- Affected lot / code info
- 127, 113, 117, 112, 131, 128, 90, 17, 35, 137, 93, 51, 95, 27, 6, 47, 98, 22, 49, 39, 103, 120, 21, 40, 70, 101, 69, 94, 89, 23, 33, 31, 30, 29, 28, 32, 100, 63, 61 Additional units are identified with site numbers: 533397, 531926, 104950, 104953, 103283, 103281, 530653, 104497
Why it was recalled
Potential for generator to lock-up. Fluoroscopy and x-ray is no longer possible, and system must be restarted to be able to continue.
Root cause (FDA determination)
Other
Action the firm took
On 10/12/06 the firm issued a letter to their consignees advising of the problem and how to avoid the problem by using specific image formats. If the system should lock up, consignees are to switch the system off and then restart. The firm will be conducting software upgrades to the devices once software is ready (estimated time for FD 10/10 is November 2006 and estimated time for FD 20/10 is January 2007). Later, the firm identified 8 additional units and issued a letter on 12/11/06 to affected customers.
Recalling firm
- Firm
- Philips Medical Systems North America Co. Phillips
- Address
- 22100 Bothell Everett Hwy, Bothell, Washington 98021-8431
Distribution
- Distribution pattern
- Nationwide.
Timeline
- Recall initiated
- 2006-10-12
- Posted by FDA
- 2006-11-14
- Terminated
- 2010-09-24
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #48924. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.