Product

Terumo Advanced Perfusion System 1; 100V -120V, 15 A (circuit breaker), 50/60 Hz (20A power source required) base; Model 801763.

FDA product code
DTQConsole, Heart-Lung Machine, Cardiopulmonary Bypass
Device class
Class 2
Medical specialty
Cardiovascular
510(k) numbers
K022947
Affected lot / code info
Serial numbers 0011 through 0039, 0041 through 0308, 0310 through 0313, 0315, 0316, 0319 through 0321 and 0324 may be affected by this issue. Serial numbers 0238 and 0282 have been corrected to date.

Why it was recalled

System 1 power supply may electrically short to the main console, disrupting normal system operation by causing unexpected re-setting or shutdown of pumps and modules, false alarms and/or overheating and failure of the central control monitor.

Root cause (FDA determination)

Other

Action the firm took

The firm corrected the complaint units at the time of the complaints in 2005. The firm issued a Device Correction Letter dated 6/6/07 notifying all users of the problem and advising them that their system will be corrected when replacement components become available.

Recalling firm

Firm
Terumo Cardiovascular Systems Corp
Address
6200 Jackson Rd, Ann Arbor, Michigan 48103-9586

Distribution

Distribution pattern
Worldwide: USA, Australia, Belgium, Canada, Dominican Republic, Egypt, Germany, Hong Kong, India, Indonesia, Japan, Malaysia, Mexico, Pakistan and Singapore.

Timeline

Recall initiated
2005-02-01
Posted by FDA
2007-06-30
Terminated
2009-07-07
Status

Source: openFDA Device Recall endpoint. Recall record ID #50175. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.