—#50175
Product
Terumo Advanced Perfusion System 1; 100V -120V, 15 A (circuit breaker), 50/60 Hz (20A power source required) base; Model 801763.
- FDA product code
- DTQ — Console, Heart-Lung Machine, Cardiopulmonary Bypass
- Device class
- Class 2
- Medical specialty
- Cardiovascular
- 510(k) numbers
- K022947
- Affected lot / code info
- Serial numbers 0011 through 0039, 0041 through 0308, 0310 through 0313, 0315, 0316, 0319 through 0321 and 0324 may be affected by this issue. Serial numbers 0238 and 0282 have been corrected to date.
Why it was recalled
System 1 power supply may electrically short to the main console, disrupting normal system operation by causing unexpected re-setting or shutdown of pumps and modules, false alarms and/or overheating and failure of the central control monitor.
Root cause (FDA determination)
Other
Action the firm took
The firm corrected the complaint units at the time of the complaints in 2005. The firm issued a Device Correction Letter dated 6/6/07 notifying all users of the problem and advising them that their system will be corrected when replacement components become available.
Recalling firm
- Firm
- Terumo Cardiovascular Systems Corp
- Address
- 6200 Jackson Rd, Ann Arbor, Michigan 48103-9586
Distribution
- Distribution pattern
- Worldwide: USA, Australia, Belgium, Canada, Dominican Republic, Egypt, Germany, Hong Kong, India, Indonesia, Japan, Malaysia, Mexico, Pakistan and Singapore.
Timeline
- Recall initiated
- 2005-02-01
- Posted by FDA
- 2007-06-30
- Terminated
- 2009-07-07
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #50175. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.