Recalls / —
—#50179
Product
Terumo Advanced Perfusion System 1, 4 inch diameter Roller Pump; Model 801040.
- FDA product code
- DTQ — Console, Heart-Lung Machine, Cardiopulmonary Bypass
- Device class
- Class 2
- Medical specialty
- Cardiovascular
- 510(k) numbers
- K022947
- Affected lot / code info
- Serial numbers 0031 through 0324 are potentially affected by the problem. Serial numbers 0092, 0128, 0152, 0223, 0226, 0237, 0246, 0260 and 0296 have been corrected.
Why it was recalled
A belt slip message may be displayed on the roller pump due to low belt tension or the presence of grease.
Root cause (FDA determination)
Other
Action the firm took
The firm is visiting consignees and making corrections as complaints are received. These corrections, to date, have been made between 04/04 and 09/06. The firm will correct other units as complaints are received and issued a Device Correction Letter dated 6/6 /07 notifying all users of the problem and appropriate responses should it occur.
Recalling firm
- Firm
- Terumo Cardiovascular Systems Corp
- Address
- 6200 Jackson Rd, Ann Arbor, Michigan 48103-9586
Distribution
- Distribution pattern
- Worldwide: USA, Belgium, Canada, Egypt, Germany, Hong Kong, India, Indonesia, Japan, Korea, Malaysia, Singapore and United Arab Emirates.
Timeline
- Recall initiated
- 2004-04-01
- Posted by FDA
- 2007-06-30
- Terminated
- 2008-06-11
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #50179. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.