FDA Device Recalls

Recalls /

#55336

Product

Shelf Pack: Bard-Parker Rib-Back Stainless Steel no.11, surgical blade. Unit Package: Contents of unopened or undamaged package guaranteed sterile; After use place product in a puncture resistant container and dispose properly. Becton Dickinson Becton Dickinson Acute Care Franklin Lakes, NJ 07417

FDA product code
GESBlade, Scalpel
Device class
Class 1
Medical specialty
General, Plastic Surgery
Affected lot / code info
Reorder number 371211; Lot Number 02028051 -Stainless Steel; Reorder number 371211; Lot Number 02028090- Stainless Steel.

Why it was recalled

Complaints received of rust on the blades of BD Bard-Parker Surgical blades.

Root cause (FDA determination)

Packaging change control

Action the firm took

Urgent recall letters were sent out 8/10/2007 via UPS second day air.

Recalling firm

Firm
Becton Dickinson & Company
Address
1 Becton Dr, Franklin Lakes, New Jersey 07417-1815

Distribution

Distribution pattern
Nationwide.

Timeline

Recall initiated
2007-08-10
Posted by FDA
2007-10-02
Terminated
2008-03-07
Status

Source: openFDA Device Recall endpoint. Recall record ID #55336. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.