Product

Baxter Colleague CX Single Channel Volumetric Infusion Pumps; Baxter Healthcare Corporation, Medication Delivery Division, Deerfield, IL 60015 U.S.A., Made in Singapore; product code 2M8161

FDA product code
FRNPump, Infusion
Device class
Class 2
Medical specialty
General Hospital
510(k) numbers
K063696
Affected lot / code info
product code 2M8161, serial numbers 14051565CC, 15034273CC, 15034492CC, 15040255CC, 15041116CC, 15041169CC, 15041170CC, 15041172CC, 15041181CC, 15041194CC, 15041199CC, 15041206CC, 15041212CC, 15041218CC, 15041241CC, 15041288CC, 15041290CC, 15041317CC, 15050020CC, 15050034CC, 15050045CC, 15050061CC, 15050080CC, 15050117CC, 15050129CC, 15060087CC, 15060217CC, 15060253CC, 15060479CC, 15060489CC, 15060549CC, 15060557CC, 15060600CC, 15060738CC, 15060790CC, 15060827CC, 15060939CC, 15060940CC, 15060968CC, 15063211CC, 15075371CC, 15075103CC

Why it was recalled

Some of the pumps were upgraded by one service technician who failed to replace prisms and batteries as required in the upgrade.

Root cause (FDA determination)

Other

Action the firm took

The affected hospitals were sent an Urgent Colleague Deployment Correction letter dated 8/15/07 via overnight mail. The hospitals were informed that some of the Colleague pumps at their facility may not have been completely upgraded to include all the necessary components required as part of Baxter's Corrective Action Plan (CAP). The hospitals were provided a list of affected pump serial numbers and were requested to remove the pumps from clinical use and notify all end users of the action. The accounts were requested to complete the enclosed customer reply form, acknowledging the receipt and understanding of the letter and fax it back to Baxter at 1-847-270-5457. The hospitals were also requested to report any malfunction or adverse event related to the Colleague Infusion Pumps to Baxter Corporate Product Surveillance at 1-800-437-5176.

Recalling firm

Firm
Baxter Healthcare Corp. Rt.
Address
120 & Wilson Rd, Round Lake, Illinois 60073

Distribution

Distribution pattern
Florida and California

Timeline

Recall initiated
2007-08-15
Posted by FDA
2007-09-19
Terminated
2008-02-21