FDA Device Recalls

Recalls /

#64427

Product

Terumo Advanced Perfusion System 1; 220V -240V, 7A (circuit breaker), 50/60 Hz (10A power source required) base; Model 801764. (Not distributed within the United States).

FDA product code
DTQConsole, Heart-Lung Machine, Cardiopulmonary Bypass
Device class
Class 2
Medical specialty
Cardiovascular
510(k) numbers
K022947
Affected lot / code info
Serial numbers 0188 through 1027.

Why it was recalled

The power supply may fail to charge the batteries due to various hardware malfunctions.

Root cause (FDA determination)

Nonconforming Material/Component

Action the firm took

Terumo notified U.S. consignees via Urgent Medical Device Correction: Safety Advisory recall letter dated 4/17/08 and international consignees were notified via email on 4/24/08 informing them of the potential problem.

Recalling firm

Firm
Terumo Cardiovascular Systems Corporation
Address
6200 Jackson Road, Ann Arbor, Michigan 48103-9586

Distribution

Distribution pattern
Worldwide distribution: USA, Australia, Belgium, Canada, Colombia, Dominican Republic, Honduras, Hong Kong, Indonesia, Japan, Jordan, Kuwait, Malaysia, Mexico, Philippines, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, United Arab Emirates, and Viet Nam.

Timeline

Recall initiated
2008-04-17
Posted by FDA
2008-12-29
Terminated
2009-07-07
Status

Source: openFDA Device Recall endpoint. Recall record ID #64427. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.

— recall: Terumo Advanced Perfusion System 1; 220V -240V, 7A (circuit breaker), 50/60 Hz (10A power source req · FDA Device Recalls