FDA Device Recalls

Recalls /

#64428

Product

Terumo Advanced Perfusion System 1 (4 inch diameter Roller Pump; Model 801040). Extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures.

FDA product code
DTQConsole, Heart-Lung Machine, Cardiopulmonary Bypass
Device class
Class 2
Medical specialty
Cardiovascular
510(k) numbers
K022947
Affected lot / code info
Serial numbers 0028 through 0111, 0113 through 0527, 0700 through 1528, 1539 through 1579, 2000 through 2392 and 2395 through 2466.

Why it was recalled

Overspeed and underspeed errors, erratic pump behavior, jerky operation at low RPMs, pump instability, pump slowdowns and pump stops due to overspeed max events may occur due to a grease leak onto the motor speed encoder disk.

Root cause (FDA determination)

Component design/selection

Action the firm took

Consignees were notified of the problem via recall letter dated 7/18/08. The firm began conducting field upgrades on the roller pumps in 07/08. For questions or concerns, contact Terumo CVS Customer Service at 1-800-521-2818.

Recalling firm

Firm
Terumo Cardiovascular Systems Corporation
Address
6200 Jackson Rd, Ann Arbor, Michigan 48103

Distribution

Distribution pattern
Worldwide Distribution including USA and countries of Australia, Belgium. Canada, Columbia, Dominican Republic, Egypt, Germany, Honduras, Hong Kong, India, Indonesia, Israel, Japan, Jordan, Kuwait, Malaysia, Mexico, Pakistan, Philippines, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey, United Arab Emirates and Viet Nam.

Timeline

Recall initiated
2008-07-18
Posted by FDA
2008-12-11
Terminated
2009-07-07
Status

Source: openFDA Device Recall endpoint. Recall record ID #64428. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.