Product

Terumo Advanced Perfusion System 1 (6 inch diameter Roller Pump; Model 801041). Extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures.

FDA product code
DTQConsole, Heart-Lung Machine, Cardiopulmonary Bypass
Device class
Class 2
Medical specialty
Cardiovascular
510(k) numbers
K022947
Affected lot / code info
Serial numbers 0030, 0032 through 0587, 0700 through 1936, 1947 through 1951 and 2000 through 2799.

Why it was recalled

Overspeed and underspeed errors, erratic pump behavior, jerky operation at low RPMs, pump instability, pump slowdowns and pump stops due to overspeed max events may occur due to a grease leak onto the motor speed encoder disk.

Root cause (FDA determination)

Component design/selection

Action the firm took

Consignees were notified of the problem via recall letter dated 7/18/08. The firm began conducting field upgrades on the roller pumps in 07/08. For questions or concerns, contact Terumo CVS Customer Service at 1-800-521-2818.

Recalling firm

Firm
Terumo Cardiovascular Systems Corporation
Address
6200 Jackson Rd, Ann Arbor, Michigan 48103

Distribution

Distribution pattern
Worldwide Distribution including USA and countries of Australia, Belgium. Canada, Columbia, Dominican Republic, Egypt, Germany, Honduras, Hong Kong, India, Indonesia, Israel, Japan, Jordan, Kuwait, Malaysia, Mexico, Pakistan, Philippines, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey, United Arab Emirates and Viet Nam.

Timeline

Recall initiated
2008-07-18
Posted by FDA
2008-12-11
Terminated
2009-07-07