Recalls / —
—#64514
Product
Philips Digital Diagnost X-Ray system
- FDA product code
- MQB — Solid State X-Ray Imager (Flat Panel/Digital Imager)
- Device class
- Class 2
- Medical specialty
- Radiology
- 510(k) numbers
- K982795
- Affected lot / code info
- Site numbers 103982, 104073, 105158, 105629, 504343, 505773. Added sites: 104303, 38280, 44807, 59667, 76907, 76909, 76974, 82641, 82652, 82657, 82724, 83072, 83073, 84712, 84862, 84863, 86072, 86987, 86988, 102030, 102100, 102489, 102713, 102717, 102719, 102787, 102905, 102911, 103109, 103165, 103274, 103437, 103438, 103439, 103453, 103893, 103894, 103941, 104465, 104902, 104903, 104904, 104937, 104938, 105415, 105621, 105623, 105625, 105668, 250094, 250470, 250471, 250875, 500002, 504546, 504989, 505038, 505347, 505350, 505351, 505352, 505605, 505684, 505944, 506184, 506343, 506643, 506645, 506826, 506839, 506848, 519148, 519149, 519150, 519568, 519739, 519904, 520297, 520298, 520299, 520385, 520687, 520698, 520766, 520808, 520811, 520828, 520836, 520978, 520984, 520986, 521004, 521089, 521110, 521412, 521539, 521660, 521668, 521893, 521894, 521995, 521997, 522213, 522215, 522244, 522281, 522360, 522361, 522362, 522364, 522367, 522398, 522475, 530411, 530412, 530415, 530455, 530516, 530611, 530696, 530716, 530779, 530787, 530789, 530791, 530792, 530796, 530800, 530806, 531773, 531825, 531833, 531858, 531975, 531994, 531995, 532138, 532141, 532167, 532168, 532169, 532171, 532206, 532211, 532278, 532364, 532421, 532422, 532815, 532816, 532817, 532818, 533244, 533317, 533319, 533338, 533426, 533468, 533534, 533571, 533769, 533915, 533916, 533920, 533949, 533950, 534082, 534279, 534435, 534539, 534570, 534622, 534718, 534800, 534801, 534836, 534866, 535146, 535334, 535381, 535383, 535549, 535626, 535773, 535774, 535786, 535933, 536002, 536039, 536043, 536062, 536064, 536128, 536129, 536441, 536448, 536614, 536913, 536939, 537163, 537184, 537455, 537753, 537755, 537762, 537840, 537918, 537948, 537950, 537974, 538139, 538154, 538166, 538304, 538437, 538478, 539680, 540173, 540199, 540474, 540477, 542194, 543376, 545097, 549011, 549725, and 557777.
Why it was recalled
Units do not comply with the requirement to provide automatic beam limitation within +/- 3 degrees of horizontal, vertical, and perpendicular to the plane of the image receptor during radiographic procedures.
Root cause (FDA determination)
Other
Action the firm took
On 7/23/07 Philips mailed a letter dated July 9, 2007 titled Electronic Product Radiation Warning. The letter advises customers of the issue and asks that customers apply the enclosed cautionary label to the device. The letter also explains that there will be a system modification to be installed. On 8/20/07 the firm sent the same letter, along with a letter dated 8/20/07 to customers thought to have either scrapped their units and the customer had previously indicated they were no longer using the unit. The 8/20/07 letter requested the customer contact Philps if they are currently using the device.
Recalling firm
- Firm
- Philips Medical Systems North America Co. Phillips
- Address
- 22100 Bothell Everett Hwy, Bothell, Washington 98021-8431
Distribution
- Distribution pattern
- Units were distributed to 1543 hospitals and medical centers throughout the US. No international distribution from this firm.
Timeline
- Recall initiated
- 2007-07-23
- Posted by FDA
- 2007-09-25
- Terminated
- 2012-09-17
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #64514. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.