Product

Argon Medical Devices, ArgoGuide Hydrophilic NiTi Guidewire, K011905, Catalog Number: 00-AG387308, .035/180cm, stiff/straight/standard, sold as a single hydrophilic guidewire in both sterile and bulk non-sterile packaging configurations. Argon Medical Devices, Inc. Athens, TX 75751

FDA product code
DQXWire, Guide, Catheter
Device class
Class 2
Medical specialty
Cardiovascular
510(k) numbers
K011905
Affected lot / code info
Lot Numbers: 99502708, and 99510894

Why it was recalled

Degradation -- Hydrophilic guidewires may exhibit degradation of blue Pebax cladding causing particulate to dislodge potentially resulting in embolism.

Root cause (FDA determination)

Process design

Action the firm took

Argon Medical sent an Urgent Product Recall letter to consignees on 11/15/2007. They requested that users immediately quarantine and return any product remaining at their facility, utilizing instruction on response card. Distributors were directed to issue the recall to their customers and return all retrieved product to Argon.

Recalling firm

Firm
Argon Medical Devices, Inc
Address
1445 Flat Creek Rd, Athens, Texas 75751-5002

Distribution

Distribution pattern
Worldwide Distribution -- USA including states of FL, KS, NV, IL, CA, NY, CO, VA, MD, NC, UT, GA, MI, NJ, TN, MA, and CT, and countries of The Netherlands, Germany, and Turkey.

Timeline

Recall initiated
2007-11-19
Posted by FDA
2008-01-31
Terminated
2008-06-05
Status