FDA Device Recalls

Recalls /

#67333

Product

CORDIS "Dura Star" 3.00 x 10 Dilatation Catheter, Catalog # 70110300, Distributed by Cordis Corporation, Miami Lakes, FL 33014

FDA product code
LOXCatheters, Transluminal Coronary Angioplasty, Percutaneous
Device class
Class 2
Medical specialty
Cardiovascular
PMA numbers
P880003
Affected lot / code info
Lot #s: 13175335 13180558 13182673 13189425 13198875 13207206 13208066 13209922 13211075 13217618 13237269 13245566 13251127 13255234 13266370 13269605 13271045 13272424 13275164 13281459 13281460 13281461 13282478 13287294 13289742 13291882 13298188 13298781 13302297 13312712 13312716 and 13313278

Why it was recalled

Slow Deflation or No Deflation

Root cause (FDA determination)

Process control

Action the firm took

Cordis initated the recall on 1/14/2008. For U.S. customers: An Urgent Medical Device Recall letter and an Acknowledgement Form was sent to the following 4 individuals in each account: Cath Lab Director, Operating Room Director, Materials Director, Risk Manager; to be followed up with direct Cordis representative contact to facilitate obtaining signature, faxing form, collecting and returning units. For non-US customers, email notification with return acknowledgement is used to notify the affiliate distributors, who then notify and reconcile the customers in those countries. All users were instructed to immediately set aside product and return it for replacement.

Recalling firm

Firm
Cordis Corporation
Address
14201 NW 60th Avenue, Miami Lakes, Florida 33014-2802

Distribution

Distribution pattern
Worldwide.

Timeline

Recall initiated
2008-01-14
Posted by FDA
2008-02-06
Terminated
2009-10-20
Status