Recalls / —
—#67878
Product
CORDIS "Fire Star" 3.00 x 10 Dilatation Catheter, Catalog # 80110300, Distributed by Cordis Corporation, Miami Lakes, FL 33014
- FDA product code
- LOX — Catheters, Transluminal Coronary Angioplasty, Percutaneous
- Device class
- Class 2
- Medical specialty
- Cardiovascular
- PMA numbers
- P880003
- Affected lot / code info
- Lot #s: 13187443 13191353 13191803 13197618 13205388 13214550 13224936 13227004 13237905 13260784 13264548 13265028 13268849 13281188 13281275 13281292 13281403 13281433 13289749 13293580 13293673 13295104 13301356 13305496 13309110 13314816 and 13323195
Why it was recalled
Slow Deflation or No Deflation
Root cause (FDA determination)
Process control
Action the firm took
Cordis initated the recall on 1/14/2008. For U.S. customers: An Urgent Medical Device Recall letter and an Acknowledgement Form was sent to the following 4 individuals in each account: Cath Lab Director, Operating Room Director, Materials Director, Risk Manager; to be followed up with direct Cordis representative contact to facilitate obtaining signature, faxing form, collecting and returning units. For non-US customers, email notification with return acknowledgement is used to notify the affiliate distributors, who then notify and reconcile the customers in those countries. All users were instructed to immediately set aside product and return it for replacement.
Recalling firm
- Firm
- Cordis Corporation
- Address
- 14201 NW 60th Avenue, Miami Lakes, Florida 33014-2802
Distribution
- Distribution pattern
- Worldwide.
Timeline
- Recall initiated
- 2008-01-14
- Posted by FDA
- 2008-02-06
- Terminated
- 2009-10-20
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #67878. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.