Recalls / —
—#67882
Product
CORDIS "Fire Star" 1.50 x 15, Single Marker Band, Dilatation Catheter, Catalog # 8011515S, Distributed by Cordis Corporation, Miami Lakes, FL 33014
- FDA product code
- LOX — Catheters, Transluminal Coronary Angioplasty, Percutaneous
- Device class
- Class 2
- Medical specialty
- Cardiovascular
- PMA numbers
- P880003
- Affected lot / code info
- Lot #s: 13205390 13206126 13214552 13215775 13237700 13239276 13252064 13278334 13281222 13281272 13281303 13281394 13281434 13292698 13293675 13294203 13294204 13294205 13294206 13301365 13303084 and 13323515
Why it was recalled
Slow Deflation or No Deflation
Root cause (FDA determination)
Process control
Action the firm took
Cordis initated the recall on 1/14/2008. For U.S. customers: An Urgent Medical Device Recall letter and an Acknowledgement Form was sent to the following 4 individuals in each account: Cath Lab Director, Operating Room Director, Materials Director, Risk Manager; to be followed up with direct Cordis representative contact to facilitate obtaining signature, faxing form, collecting and returning units. For non-US customers, email notification with return acknowledgement is used to notify the affiliate distributors, who then notify and reconcile the customers in those countries. All users were instructed to immediately set aside product and return it for replacement.
Recalling firm
- Firm
- Cordis Corporation
- Address
- 14201 NW 60th Avenue, Miami Lakes, Florida 33014-2802
Distribution
- Distribution pattern
- Worldwide.
Timeline
- Recall initiated
- 2008-01-14
- Posted by FDA
- 2008-02-06
- Terminated
- 2009-10-20
- Status
- —
Source: openFDA Device Recall endpoint. Recall record ID #67882. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.