FDA Device Recalls

Recalls /

#67898

Product

CORDIS "Fire Star" 2.25 x 20 Dilatation Catheter, Catalog # 80120225, Distributed by Cordis Corporation, Miami Lakes, FL 33014

FDA product code
LOXCatheters, Transluminal Coronary Angioplasty, Percutaneous
Device class
Class 2
Medical specialty
Cardiovascular
PMA numbers
P880003
Affected lot / code info
Lot #s: 13211694 13221103 13233515 13243868 13249524 13254624 13261566 13264098 13276282 13279281 13280402 13281428 13286584 13289759 13294553 13298193 13321819 and 13325806

Why it was recalled

Slow Deflation or No Deflation

Root cause (FDA determination)

Process control

Action the firm took

Cordis initated the recall on 1/14/2008. For U.S. customers: An Urgent Medical Device Recall letter and an Acknowledgement Form was sent to the following 4 individuals in each account: Cath Lab Director, Operating Room Director, Materials Director, Risk Manager; to be followed up with direct Cordis representative contact to facilitate obtaining signature, faxing form, collecting and returning units. For non-US customers, email notification with return acknowledgement is used to notify the affiliate distributors, who then notify and reconcile the customers in those countries. All users were instructed to immediately set aside product and return it for replacement.

Recalling firm

Firm
Cordis Corporation
Address
14201 NW 60th Avenue, Miami Lakes, Florida 33014-2802

Distribution

Distribution pattern
Worldwide.

Timeline

Recall initiated
2008-01-14
Posted by FDA
2008-02-06
Terminated
2009-10-20
Status

Source: openFDA Device Recall endpoint. Recall record ID #67898. The FDA issues recall classifications as health-hazard assessments, not legal findings; for legal claims consult a licensed attorney.